User Guide
Page 3 - 10
Genesis WDG-1200/1210
Insitu Oxygen Probes
3.3 Electrical Connections
The customer is responsible for providing all switch cabinets, distribution boxes,
fuses and other components for electrical installation as well as the mains power
supply connections.
NOTE
A suitable switch or circuit breaker must be provided close to the probe to
allow the electrical power supply to be disconnected.
Mains Power Supply
Mains power connections must be made for:
•
the oxygen probe
•
the optional remote calibration unit (where applicable)
•
any additional equipment (e.g., chart recorder, etc.)
Remote Calibration Unit (optional)
Signal connections must be made for:
•
Automatic Trigger
•
Pressure Signals
WARNING
The automatic trigger circuit is not isolated from internal DC circuits
and provides only for Basic insulation. Circuits connected to this
set of terminals must be connected only to appropriate components
and properly insulated to provide at least an additional level of Basic
insulation to assure user safety.
NOTE
Both the WDG-1200 Insitu and the WDG-1210 Insitu are designed to work
in conjunction with the 1200/1210 Remote Calibration Gas Unit.