User Guide
Page 3 - 22
Genesis WDG-1200/1210
Insitu Oxygen Probes
3.4 Pre-operational Checks - Final Checks
Check that a
blanking plug
has been securely fitted to the
calibration gas
connector
on the probe.
NOTE
If the blanking plug is not fitted, air leaking into the probe via the connector
can cause measurement errors. In a pressurised flue, gases venting to
atmosphere through the connector could cause corrosion of the test gas
tube. In a negative pressure flue, air leakage can cause high O
2
reading
errors.