User Guide
Page 4 - 25
Genesis WDG-1200/1210
Insitu Oxygen Probes
The ‘Status and Calibration Start’ register 20 is used both to trigger a calibration
and to monitor the instrument status. The meaning of each bit is as shown in
the table below:
Bit
Meaning
0 (LSB)
Start Calibration.
Write “1” to start calibration
1
Zero Calibration acknowledged
2
Span Calibration acknowledged
3 to 9
Not used
10
Failed in Calibration
11
Cold Junction Error
12
Cell too hot
13
Zero Coefficient out of range
14
Span Calibration out of range
15 (MSB)
Temperature out of range
(± 5 degrees)