User Guide
Page 5 - 4
Genesis WDG-1200/1210
Insitu Oxygen Probes
Fault Description
Key to Elements column of each fault:
Probe
Fault is related to probe
Wiring
Fault is related to wiring of system
Mains
Fault is related to mains power supply
ACU
Fault is related to Remote Calibration Unit
NOTE
For all actions referring to the Remote Calibration Unit refer to the relevant
operating manual.
(a) Heater temperature out of control band 00.02.
The probe temperature has not increased/decreased or is too high during the
warm up period.
CAUSE
ELEMENT
ACTION
1. Heater not connected.
Probe
Check probe wiring.
2. Heater fuse blown.
Probe
Replace heater fuse and check
wiring.
3. Heater failed.
Probe
Replace probe heater.
4. Thermocouple failure.
Probe
See faults under
Thermocouple fault
and heater above maximum limit
,
identification number 40.00.
5. Cold junction sensor
failed.
Probe
See faults under
Cold junction
temperature fault
, identification
number 600.0.
6. Signal noise.
Wiring
Re-route probe signal cable away
from power cables.
7. Mains voltage low.
Mains
Obtain suitable analyzer supply.
8. Electronic failure.
Probe
Contact Service Engineer