Page 4 - 16
Genesis WDG-1200/1210
Insitu Oxygen Probes
Probe type (SySt
↵
Prty)
To select the probe type:
Press
Enter
Select a value using the
UP
and
DOWN
keys
Press
Enter
again.
NOTE: The default setting corresponds to the Model
1200/1210 probe. Do not change this setting.
OPTIONS
Set probe type
RANGE
673 to 1172
DEFAULT
973
Baud setting (SySt
↵
bAud)
To set the communications speed:
Press
Enter
Select one of the 4 options using the
UP
and
DOWN
keys
Press
Enter
again.
NOTE: 96np is 9600 baud with no parity bit.
OPTIONS
[1200][4800][96np][9600]
RANGE
NOT APPLICABLE
DEFAULT
9600
Negative values (SySt
↵
nEg)
Negative values are not normally displayed but are
sometimes required for diagnostic purposes. The
system can be set to display negative values by
pressing
Enter
, selecting
YES
using the
UP
key, and
then pressing
Enter
again. Selecting
NO
using the
DOWN
key returns the display to normal.
OPTIONS
[NO][YES]
RANGE
NOT APPLICABLE
DEFAULT
NO
Probe address (SySt
↵
PrAd)
To set the probe address for the Modbus protocol:
Press
Enter
Select a value using the
UP
and
DOWN
keys
Press
Enter
again.
OPTIONS
Set probe address
RANGE
0 to 255
DEFAULT
11
Revision number of software (SySt
↵
rXXX)
The software revision number used in the analyzer
can be displayed by pressing
Enter
. Example shown
is release 1.04.
OPTIONS
Display software revision number
RANGE
NOT APPLICABLE
DEFAULT
NOT APPLICABLE
Summary of Contents for LAND Genesis WDG-1200
Page 4: ...User Guide Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Page 6: ...User Guide Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Page 14: ...User Guide Page 2 4 Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Page 68: ...User Guide Page 4 32 Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Page 94: ......
Page 95: ......