Alcatel Vacuum Technology France --- IPUP A100 Instruction manual
Use of the monitoring system for
pumping operation
PUMP START UP
Start up the pump by setting 24 VDC on the Amat SPI connector (pins
1---2)
(see B 60)
with 24 VDC.
The following operations will occur automatically:
--- water valves open (IPUP A100P version only),
--- IPUP A100 pump starts,
--- data capture and sensor processing,
--- purge gas valve opens (IPUP A100P version only), if option CMD
PURGE was selected (see Neutral gas purge control).
Then adjust the gas flow rate using the pressure regulator (as required
by customer application)
(see C 60)
.
At the same time the operating parameters are displayed.
* 19H REM N2
PRESSURE
ADP TEMP
500 MBAR
110° C
To configure the pump parameters, use the
SET
key
(see C 50 and
C 51)
.
Only installed and configured parameters are displayed.
l
Do not move a pump when in operation.
Note : During pump start up, the temperature is not monitored (no
warning, no hazard) until the pump has reached the programmed
temperature value (P version only).