D 80
1 / 2
Alcatel Vacuum Technology France --- IPUP A100 Instruction manual
Fault message display --- Facilities
problem
IPUP A100P WITH HAND HELD
DISPLAY MODULE
Pump is running and displays a
warning
Warning
then
INCIDENT
CAUSE
TROUBLESHOOTING
NO
No purge flow
l
Check nitrogen supply.
NO
Insufficient purge flow
l
Check the setting of
nitrogen flow.
Other problem
l
Call your Alcatel Service
Representative.
Hazard
D13: N2 FLOW
W13: N2 FLOW
NO
Incorrect N2 warning level
setting
l
Check that warning level is
correctly set.
G
B
00
55
1