Saving and loading of pump
configuration (with monitoring
system)
For pump installation or replacement, the user may have to copy the
configuration (set of all parameters and operation setting programed
by the user) from one pump to another.
Loading and saving the parameters with the pump in
operation may result in a software conflict.
Perform these operations when the pump is switched on but
with pumping stopped.
Saving the configuration
The pump configuration is stored in the monitoring system.
Enter in the menu by pressing
SET
.
Press the
+
key repeatedly to have
acces to the MANAGEMENT menu
.
Validate with the
ENTER
key
.
With the
+
key, move through to
CONFIG. SAVE menu
.
Validate with the
ENTER
key
.
The pump configuration is then transfered into the hand held dislay
module memory.
Return to main menu and exit the setting mode by pressing
SET
.
Disconnect the hand held display module.
Remote loading
Connect the hand held dislay module onto the pump to be configured.
Load the configuration previously saved in the hand held display
module into the pump by selecting the CONFIG LOAD menu from
MANAGEMENT menu.
The former configuration is automatically reset.
Loading parameters may result in creating software conflicts.
After loading, configure manually the following parameters
(temperature unit, pressure unit, serial link, time and date).