11. Programming With Programming-Menus
77
charge automatically ended. Then only will the network icon. Batteries are
loaded correctly in all cases. The external charge adapter from rechargeable
battery set ZA2690-AS cannot be used with this device.
11.8
The locking and calibration menu (option KL)
In the
"Locking and calibration" menu
you can
lock the right-of-access to certain menus and to
certain functions. Here you can also see the se-
rial numbers and calibration data for the device
itself and for any sensors attached. With option
KL it is possible not only to correct the sensor at
several points in the connector itself but also to
manage the associated calibration data.
The right-of-access to this and other menus and
to certain key functions can be stipulated in detail
and protected by password by means of the pa-
rameters "Menu" and "Fct".
Menu Locking the menus
0
none
1
+ calibration menu, except password
2
+ programming menus, except recording to memory and output from memory
3
+ recording to memory and output from memory
4
+ wizard menus
5
+ measuring menus, except user menu U1
Fct Locking the functions
and keys
0
none
1
+ data input, switching on and off
PROG, ON, OFF, ZERO, ADJ
2
+ clearing measured data
CMEM, CLR, CLRA
3
+ start / stop / output measuring operation
START/STOP, MANU, ARRAY, PRINT
4
+ function selection, measuring point selection
PROG, F«, M«
No password, locking with new password:
Password : -
-
-
-
Locked with password, enter correct password :
Password : ****
Select locking level, menu, and function :
Locking mode : Menu: 0 Fct: 0
The device type (with version and serial number) and the sensors (with order
number and serial number) are displayed. With option KL you can enter the date
of the next calibration and the calibration interval in months. If “Calibration mes-
sage” is activated then, as soon as the next calibration is due, a message to this
effect will appear when the device is switched on.
* LOCKING-CALIBRATION *
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Password:
****
Locking level: Menu: 0 Fct: 0
Device:
2890-9 6.22
Serial number:
04020123
Next calibration:
01.12.05
Signal for calibration:
Ø
Sensor:
Channel: 00
Type:
FHA646-6
Serial number:
04020123
Next calibration::
01.02.06
Calibr. interval:
12 Month
PRINT ESC