11. Programming With Programming-Menus
71
11.5.3
Baud Rate, Data Format
All interface modules are factory-set and programmed to 9600 baud. To avoid
unnecessary problems when networking several devices the baud rate should
not be modified but the computer should be set up accordingly. If this is not
possible the values 1200, 2400, 4800, 9600bd or 57.6, 115.2 kbd can be entered
in the function
Baud Rate
(please consider the max. baud rate of the interface
module!). The baud rate setting will be stored in the EEPROM of the interface
module and will then be valid for use with all other ALMEMO devices.
Function
Baud Rate
:
BaudRate: 9600bd
Data Format:
unchangeable 8 data bits, no parity, 1 stop bit
11.5.4
Language
The language of the function legend and the outputs can be set to German,
French or English. The soft keys are international code and are not meant to be
altered:
Language selection in function
Language
s. 9.5:
Language: English
11.5.5
Lighting and Contrast
The background lighting of the display can be switched on or off in 3 levels in
the selection menus by using the key
<* ON>
or in the function
Lighting
of the
device configuration (attention, the power consumption is fife times higher in
level 3!). If the lighting is switched on when a mains supply adapter is not con-
nected, the lighting will switch off in a configurable lighting time after the last key
was operated (pause), and it will switch on again as soon as any key is operated.
With the function
Contrast
the contrast of the display can be set to 10 different
positions.
Switching on the lighting, level 1 to 3:
Lighting level: 2
Switching off the lighting, level 0:
Lighting level: 0
Entering the lighting time from 20sec to 10 min:
Lighting time: 20s
If the
lighting is switched on
,
the status line provides the symbol:
*
Lighting on
If it is temporarily switched off, lights up:
*
Pause
Switch-on
without
function by using the key:
<ESC>
Setting the contrast (10 to 100%) s. 9.5:
Contrast: 50%
11.5.6
Atmospheric Pressure
For compensation of various sensors the atmospheric pressure can be entered
(s. 10.2.6). If it is measured, it will also be indicated in this function:
Entering the atmospheric pressure in function:
Atm. Pressure 1013mb