10. Measuring using Measuring-Menus
47
10.7
User Menus
When studying the measurement menus you may have noticed that the display
of the meas. value and the combination of functions does not always match your
applications in an optimum way. Therefore, you can freely configure the three
user menus
U1
to
U3
, in addition to the standard measuring menus using the
software ALMEMO
®
Control. You can use the following list of functions to indi-
vidually place the functions you require, in any sequence, on your display up to
the available space of 13 lines. There is no limit in the number of measuring
points, as it is at the ALMEMO
®
2590-9. In addition to the meas. functions that
have been described before, it is possible to use times and cycles for the pro-
cess control (s. 11.1.) and most of the sensor programming functions.
10.7.1
Functions
Functions:
Display:
Keys:
Command
:
Meas.value small
00: 234.5°C Temperature
ZERO ADJ
o 15
Meas. value medium
3 lines
00: 1234.5 °C
ZERO
ADJ
o 16
Meas. value large
7 lines
00:
Temperature
°C
1234.5
ZERO
ADJ
o 17
Meas. value bar chart
2 lines
|¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹¹¹¹|¹¹¹¹¹¹¹¹¹¸¹¹¹¹¹¹®®®¸®®®®®®®®®¸®®®®®®®®®¸®®®®®®®®®|
5.0 S220 mls 15.00
o 34
Limit value Max
(s. 11.3.5)
Limit Max: 1234.5°C
OFF ON
o 00
Limit value Min:
Limit Min: -0123.4°C
OFF ON
o 01
Base value
(s. 11.3.6)
Base: ------
OFF ON
o 02
Factor:
Factor: 1.12345
OFF ON
o 03
Exponent
Exponent 0
OFF ON
o 48
Zero point corr.
(s. 11.3.7)
Zero correction: ------
OFF ON
o 04
Slope correction:
Slope correction: ------
OFF ON
o 05
Analog start
(s. 11.4.4)
Analog-start: 0.0°C
OFF ON
o 06
Analog end:
Analog-end: 100.0°C
OFF ON
o 07
Range
(s. 11.3.9)
Range: NiCr
o 08
Maximum value
(s. 10.1.2)
Maximum value: 1122.3°C
CLR CLRA
o 09
Minimum value:
Minimum value: 19.3°C
CLR CLRA
o 10
Average value
(s. 10.4.5)
Average value: ------
CLR CLRA
o 11
Cycle
(s. 11.1.2)
Cycle: 00:00:00Un
CLR FORM
o 12
Time, date
(s. 11.1.1)
Time:12:34:56 Date:01.02.00
CLR
o 14
Averaging mode
(s. 10.4.2)
Averaging mode: CONT
CLR
o 18
Conv. rate:
(s. 11.1.3)
Conv.rate: 10M/s Cont: -
OFF ON
o 19
Cycle Timer:
(s. 10.3.2)
Cycle-timer: 00:00:00Un
CLR FORM
o 20
Averaging Counts
(s. 10.4.3)
Counts: 00000.
o 22