11. Programming With Programming-Menus
73
11.6.1
Data cable
The serial interface can be used to output cyclic data logs, all function values of
the measuring menus, as well as the whole programming of the sensors and
the device to a computer. The ALMEMO® data cables and the connection to the
instruments are described in the manual section 5.2. Other modules for net-
working the instruments follow in the manual section 5.3.
All available interface modules are connected to the socket A1
(2)
, exception:
network cables ZA1999-NK to network a further instrum. are connected to A2.
Under the prevailing socket the menu indicates:
Outout Socket A1:
DK Data cable
Variant 0: Standard interface always active
0: RS232
The baud rate is also stored in the cable connector:
Baudrate: 9600 Bd
11.6.2
Relais Trigger modules
Whereas, for the purposes of addressing peripheral devices via the relay and
trigger input (see Manual 5.1.2/3), V5 modules (ZA 1000-EAK, ZA 8000-RTA)
provide only one function variant (see Manual
6.6.4), the new V6 relay trigger analog adapter
ZA 8006-RTA3 offers up to 10 relays or options
with 2 of these as trigger inputs or up to 4 as
analog outputs. All elements can be configured
individually for these functional variants. Old
output cables can be reconfigured for V6 func-
tions by means of ALMEMO
®
Control. These
modules can be connected equally well to out-
put socket A2 or output socket A1
(2)
. To ensure
that all elements are addressed, each of these
sockets has been assigned 10 port addresses.
Socket Connection
Port addresses
A1
V6 output modules at socket A1
10..19
A2
V6 output modules at socket A2
20..29
In the
"Output modules "
menu the elements of the output modules can be
individually selected and functions programmed as follows (see Manual 6.10.9)
:
First
select the port
by pressing :
<P>
:
▲
or
▼
e.g. port 0, socket A2 (port address 20) :
Port: 20
The element concerned is recognized :
Relay :
Relay type = NO (normally open) :
Relay : NO
Relay type = NC (normally closed) :
Relay : NC
Relay type = changeover :
Relay : Changeover
The relay switching mode can be configured to the following
variants;
see 9.5:
* OUTPU MODULES *
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
ZA 8006-RTA3 Socket: A2
Port: 0
Adr.: 20
Relais: normally open 0.5A
2: Relais driven externally
Zustand: activ open
Watchdog: Ø Error
ON OFF P PRINT ESC