11. Programming With Programming-Menus
58
Clear memory
Select function
Memory free
(s. 9.4):
Memory free:
384.5
kB
To clear the memory, press the key:
<CMEM>
Full memory capacity is indicated in function:
Memory free: 512.0kB
Cancel by using the key:
<ESC>
11.3
Sensor Programming
As all ALMEMO® instruments contain the whole sensor programming stored in
the ALMEMO® connector plug, the user, usually, does not need to perform any
programming. Only if, for example, sensor errors must be corrected or existing
sensors must be scaled or limit values need to be specified the comprehensive
programming options have to be used.
If the corresponding sensor connector is con-
nected all parameters of a channel can be
checked and entered or changed via keypad in
the menu
SENSOR PROGRAMMING
. It must be
considered that standard sensors are, by a
locking mode, protected against unintentional
modification and that the locking level must first
be reduced before desired changes can be
performed (s. 11.3.4). The functions are only
selectable as far as the locking mode allows,
the rest appears in grey.
Output of the sensor programming of all active measuring
points (command P15 s. man. 6.2.3) with key:
<PRINT>
11.3.1
Selecting the Input Channel
To query or to program the parameters of a sensor, it is first necessary to select
the menu
SENSOR PROGRAMMING
and then to set the required input channel by
using the key
or
. For this, only connected sensors and activated
channels will be considered. To be able to activate new channels, the key
<MALL>
can be used to enable the selection of
all
channels. The key
<MACT>
can be
used to reduce the selection back to the
active
channels only. The correspond-
ing connector number will be displayed with each input channel.
Menu
SENSOR PROGRAMMING
:
Indication of connector number and channel:
Connector:0 Channel:00
Selecting the next input channel using the key:
Selecting the previous input channel using the key:
Enabling the selection of all available channels:
<MALL>
Reducing the selection to all active channels:
<MACT>
* SENSOR PROGRAMMING *
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Connector: 0
Channel: 00
Comment:
Temperatur
Averaging mode:
CONT
Locking mode:
5
7 Limit max:
35.0 °C
7 Limit min:
-----
5 Base:
-----
5 Factor:
-----
5 Exponent:
0
4 Zero correct.:
-----
4 Slope correct.:
-----
2 Dimension:
°C
1 Range:
NiCr
MALL M PRINT ESC