Agilent U1840A Medalist sj5000 Hardware Manual
6
Auto Width Adjust Conveyor Setup
Linear Mechanical Setup
The tools required for this check are a 2 mm Allen key and an
engineers steel ruler.
In order to achieve mechanical accuracy with the AWA
conveyor, carry out the following setup check and make the
necessary adjustments.
1
Use the steel ruler to measure the distance from the inside of
the linear rail to the outside of the black lip where the outer
clamp edge ends. This distance must be less than or equal to
4.75 mm along the entire the length of the rail as shown
below.
2
If the measurement exceeds 4.75 mm, proceed with the
mechanical setup described in
3
If the measurement is acceptable (less than or equal to
4.75 mm), proceed to