Setting Up Barcode Readers
10
Agilent U1840A Medalist sj5000 Hardware Manual
10-5
Set Up Multiple Barcode Readers
Each barcode reader is identified by the header that it sends to
the sj5000 software upon successful barcode reading. In the
standard MS-4 barcode setup file, the header is set to ‘0’ by
default. If more than one barcode reader is connected, the
header will have to be incremented as described below.
Make sure all barcode readers are properly connected to the
USB ports before proceeding.
1
Launch the Microscan ESP software.
2
Select the barcode reader model (
MS-4
) and click
OK
.
3
When prompted to connect to the MS-4 barcode reader
immediately, click
No
.
4
Open the standard MS-4 ESP setup file
C:\Installation\Barcode\MS4-1.esp.
5
Select
RS232
protocol.
6
Select the COM port for each barcode reader that is
connected and click
Connect
.
7
Click
Output Format
.
8
When asked whether to synchronize with the reader, click
Receive
to receive existing data output format into the screen.
Summary of Contents for U1840A Medalist sj5000
Page 15: ...Safety and Regulatory Information 1 Agilent U1840A Medalist sj5000 Hardware Manual 1 9...
Page 50: ...3 12 Agilent U1840A Medalist sj5000 Hardware Manual 3 Installation...
Page 100: ...7 18 Agilent U1840A Medalist sj5000 Hardware Manual 7 Camera and Lighting Calibration...
Page 134: ...10 18 Agilent U1840A Medalist sj5000 Hardware Manual 10 Setting Up Barcode Readers...
Page 176: ...11 42 Agilent U1840A Medalist sj5000 Hardware Manual 11 Troubleshooting...
Page 190: ...A 4 Agilent U1840A Medalist sj5000 Hardware Manual A System Specifications...
Page 192: ...B 2 Agilent U1840A Medalist sj5000 Hardware Manual B Reference Documents...