Camera and Lighting Calibration
7
Agilent U1840A Medalist sj5000 Hardware Manual
7-7
5
When the gantry has stopped moving, from the
Tools
menu
click the
Interface
folder and click the
System Information
icon.
6
Verify the values for pixel size (19 microns ±1 micron for the
sj5000) and field of view in the System Information window.
These may be different for certain applications.
7
Click
Close
and then close the
Tools
menu.