96
Installation and Verification Manual
3
Performance Tests
Optical Receiver Sensitivity, Optical Power and Frequency Measurement
Specifications
Min. sensitivity
(Note1):
52/155 Mb/s
622 Mb/s
2.5 Gb/s
2.66 Gb/s
9.95 Gb/s ;
10.71 Gb/s ;
–34 dBm
–28 dBm
–28 dBm
–26 dBm
–18 dBm (version 1)
–14 dBm (version 2)
–16 dBm (version 1)
–14 dBm (version 2)
Max. input power
(Note1)
:
52/155 Mb/s
622 Mb/s
2.5 Gb/s
2.66 Gb/s
9.95 Gb/s & 10.71 Gb/s;
–10 dBm
–9 dBm
–9 dBm
–9 dBm
–10 dBm (version 1)
–1 dBm (version 2)
Note 1;
52 Mb/s – 2.66 Gb/s; for BER = 1 x 10 –10 (input signal extinction ratio >= 8.2
dB).
9.95 Gb/s & 10.71 Gb/s; for BER = 1 x 10 –12 (input signal extinction ratio >= 8.2
dB).
Check the stated Optical 10G Output Power on the instrument panel to
determine if the instrument has version 1 or version 2 optics fitted;
Version 1 has a 10G Optical Ouput Power of +1 dBm Max.
Version 2 has a 10G Optical Ouput Power of –1 dBm Max
panther3_iv.book Page 96 Wednesday, January 15, 2003 12:03 PM