
128
Installation and Verification Manual
3
Performance Tests
1
Recall the OmniBER OTN’s default setting as shown on
2
Press
<Menu>
, select
Tx/Rx > Transmitter Settings
and set
the
Signal Mode
to
PDH/DSn
. Now set the
Signal Rate
to
DS3
.
34M Unbalanced Negative Pulse
To test the DSX-3 output pulse mask
panther3_iv.book Page 128 Wednesday, January 15, 2003 12:03 PM