Performance Tests
3
Installation and Verification Manual
117
5
Check the DSX-1 pulse amplitudes are between 2.4-3.6V pk.
6
Set the instrument Output Level to DS-1-LO and check that
the Pulse Output Amplitude decreases by approximately 20%
and pulse shape has slower rise/fall edges.
DSX-1 Positive Pulse
DSX-1 Negative Pulse
panther3_iv.book Page 117 Wednesday, January 15, 2003 12:03 PM