Installation and Verification Manual
3
Performance Tests
3
Connect the OmniBER OTN’s
DS1 OUT
port to the
Oscilloscope using the E2621A Termination Adapter Probe,
set probe switch to
100
Ω.
4
Load the
DS1_ANSI.msk
test mask and run the mask test. Check the
Positive and Negative pulses are within the mask (see examples).
Bal/Unbal
Adapter
54810A Oscilloscope
OmniBER OTN
CH1
E2621A
DS1 Balanced O/P
Set
to
100
Ω
Bantam to Bantam Cable
DS1 O/P
N O T E
To test Negative pulse set the Scope
Mask Test Set-up
panel to
Invert
Mask.