Magnis and Magnis Dx NGS Prep System User Guide
44
7
When prompted on the touchscreen for a tip box and tip, open the instrument door, add a new
tip box (with lid off) to the indicated platform, and add a tip to the micropipettor at the indicated
barrel position. Close the door, then press
Next
to continue.
8
When prompted on the touchscreen to remove the tip box, open the instrument door and
remove the tip box. Close the door, then press
Next
to continue.
The auto-teaching process continues until completion.
Include teach-point verification in the IHC
The system can include teach-point verification as part of the initial instrument health check (IHC)
that is performed after powering up the instrument. If the system detects an error during
teach-point verification, the corrective action is to run auto-teaching.
1
From the Home screen, press
Settings
.
The
2
Press
System Settings
.
The
3
Press
Other Settings
.
The
opens.
Figure 27
Other Settings screen
4
Mark the check box labeled
Verify teach points during Instrument Health Check
.
5
Press
Close
to save your changes.