Magnis and Magnis Dx NGS Prep System User Guide
27
Programming System Settings
Set the chiller temperature
When the system is running a protocol, the temperature of the chiller is dictated by the protocol
parameters. During setup of a protocol run, however, the chiller is pre-cooled to the temperature
specified on the Chiller Setting screen. The temperature is set to 12°C by default, but
temperatures from 4°C to 12°C are permitted.
Only users with Advanced access are permitted to set the chiller temperature.
1
From the Home screen, press
Settings
.
The
2
Press
System Settings
.
The
3
Press
Chiller Setting
.
The
Figure 12
Chiller Setting screen
4
On the Chiller Setting screen, enter the temperature for the chiller (°C) into the provided field.
5
Press
Close
to save your changes.
Set the time and date
The times and dates recorded in the system’s log files are based on the time and date settings for
the system.
1
From the Home screen, press
Settings
.
The
2
Press
System Settings
.
The
3
Press
Date & time Settings
.