![Agilent Technologies Magnis Dx NGS Prep System User Manual Download Page 18](http://html1.mh-extra.com/html/agilent-technologies/magnis-dx-ngs-prep-system/magnis-dx-ngs-prep-system_user-manual_2867773018.webp)
Magnis and Magnis Dx NGS Prep System User Guide
18
Instrument Status Indicator Lights
You can quickly and easily check the status of the instrument based on the color of the LED
indicator lights that illuminate the entire plate-filling area.
Table 5
Colors and descriptions of the LED indicator lights in the plate filling area
LED Color
Instrument Status
Description
White
Ready
The lights are illuminated white when the system is idle but the door is
open, when the system is performing auto-teaching or teach-point
verification, and when a user is setting up a protocol run.
Blue
Ready
The lights are illuminated blue anytime the system is idle and the door
is closed, including at completion of a protocol run. The lights are also
blue when the system is performing diagnostic tests.
Green
Running
The lights are illuminated green when the system is running a protocol.
Red
Error
The lights are illuminated red when the system has encountered an
error. Check the touchscreen for an error message containing further
details.
WARNING
During UV decontamination, the indicator lights are off and the
instrument deck is illuminated by the UV light. Do not look directly at the
UV light.