84
3000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
46
Press the
[Run/Stop]
front panel to begin repetitive acquisition again.
47
Press the
[Acquire]
front panel key.
48
Press the
Segmented
softkey; then turn the Entry knob to select
500
as
the number of segments to capture.
49
Press the
Segmented
softkey to begin a Segmented Memory acquisition.
50
Press the
Current Segment
softkey; then turn the Entry knob to review all
500 captured segments.
51
Set the
Current Segment
to
500
and note the time- tag of the last captured
segment.
Segmented Memory optimizes oscilloscope acquisition memory by only
capturing important segments of a waveform based on the trigger
condition and timebase setting. In this example, we have selectively
captured over 4 seconds of total acquisition time as shown in
Note that we can also view the decoded SPI data in the “lister” format,
and we can also perform Search & Navigation on the segments.
Figure 59 Using the scope’s Segmented Memory acquisition to selectively capture more
SPI traffic.