3000 X-Series Oscilloscopes Advanced Training Guide
3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
100 ns wide. Let’s set up the scope to uniquely trigger on the positive
runts that are < 200 ns wide. In other words, let’s isolate the smaller of
these two runts.
13
Press the
Qualifier
softkey; and then select “
<
”.
14
Press the
Time
softkey; then turn the Entry knob to select
< 200.0 ns
.
Your scope’s display should look similar to
showing a positive
runt pulse that is approximately 100 ns wide near center- screen. In
addition to triggering on runt pulse conditions, the Agilent 3000 X- Series
oscilloscopes can also perform
Search & Navigation
to find multiple runt
pulse conditions, regardless of the specific trigger setup condition. Let’s
now capture a longer stream of pulses and perform an automatic search
operation to find all runts.
15
Set the scope’s timebase to
100.0 µs/div
.
16
Press
[Run/Stop]
to stop repetitive acquisitions.
17
Press the
[Search]
front panel key.
18
Press the
Search Edge
softkey; then turn the Entry knob to select
Runt
.
19
Press the
Settings
softkey to observe the default runt search conditions.
Figure 38 Triggering on a positive runt pulse that is < 200 ns wide.