
Oscilloscope Familiarization Labs
2
3000 X-Series Oscilloscopes Advanced Training Guide
37
15
Press
Statistics
.
16
Press
Reset Statistics
to clear the statistics from the previous test.
17
Press the
(Back) key.
18
Press
Setup
.
19
Press
On Error – Stop
two times.
20
Press
[Run/Stop]
to begin a new “Stop- on- Error” test.
In addition to stopping acquisitions when an error is detected as shown in
, you can also save a waveform, save an image, print, or perform
specific measurements when an error is detected. Using the “Run Until”
selection, you can also set up mask testing to run for a specific number of
tests, minimum time, or minimum Sigma test criteria. Note that in
addition to creating a pass/fail mask using the auto- mask creation feature
as demonstrated in this lab, you can also import a mask based on industry
specifications, such as an eye- diagram mask, from a USB memory device.
These types of imported masks can be easily created within a standard
text editor, such as NotePad on your PC.
Figure 21 Establishing specific test criteria.
Back