3000 X-Series Oscilloscopes Advanced Training Guide
3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
Lab #13: Triggering on a Qualified Burst using “Edge then Edge” Trigger
In this lab you will learn how to use the scope’s “Edge then Edge” trigger
mode. This trigger mode is sometimes referred to as “qualified A then B”
triggering. “Edge then Edge” triggering allows you to qualify triggering on
an edge of any channel, then delay arming of the final trigger condition by
both time and event count of an edge of any channel.
1
Connect the channel- 1 probe to the Demo 1 terminal and ground.
2
Connect the channel- 2 probe to the Demo 2 terminal and ground.
3
Press
[Default Setup]
on the scope’s front panel.
4
Press
[Help]
; then press the
Training Signals
softkey.
5
Using the Entry knob, select the
Edge then Edge
signal; then press the
Output
softkey to turn it on.
6
Set channel- 1 to
1.0 V/div
.
7
Press the
[2]
front panel key to turn on channel- 2.
8
Set channel- 2 to
1.0 V/div
.
9
Set channel- 2 offset/position to approximately
2.0 V
.
10
Push the trigger level knob to automatically set the trigger level on both
channels to approximately
50%
.
11
Set the scope’s timebase to
1.0 µs/div
.
Figure 44 Capturing a complex burst signal while triggering on a synchronization signal.