Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page
7
of 14
Document No.
699910002
Revision
D
Figure 7: Front Panel Display
13) Dress the RS-232 cable through the leak detector to the front panel display.
14) Remove the M6 screw and washer from the front panel display and detach the display (Figure 7: Front Panel
Display).
15) Connect the RS-232 cable end labeled J300-2 to P300-COM2 on the front panel display board using its two
slotted screws (Figure 8: RS232 Cable).
Figure 8: RS232 Cable
16) Reattach the display by inserting the tabs into their slotted positions and secure using the M6 screw and washer.
17)
Attach the front cover and secure to the frame using existing
hardware.