Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page
9
of 14
Document No.
699910002
Revision
D
Section Two: Installation on G8610, G8611, or G8612 Series Leak Detector
Equipment Required
Kit Includes
Slotted Screw Driver
Wireless Base PCB
Metric Allen Wrench Set
Wireless Remote Label
M3 Philips Head Screw Driver
RS232 Cable
Adjustable Wrench
Antenna
M3 Screw (Qty. 4)
Installation Procedure
For clarity, some items have been omitted from views.
WARNING
Disconnect power from the unit before performing any maintenance
procedure that requires physically disconnecting any part of the system.
1) Turn off the power switch located on the back of the unit and unplug.
2) Wait 30 seconds for the high voltage to dissipate.
Figure 1: Orientation
3) Using slotted screw driver, unfasten the four screws holding the front, rear and right side covers (Figure 1:
Covers) and detach. The left side cover is secured with hex-head screws.
Front
Right Side
Left Side (Not Shown)
Rear (Not Shown)