Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page
14
of 14
Document No.
699910002
Revision
D
Figure 9: Label
15) Attach the label to the rear cover in location shown (Figure 9: Label). This new label will help service identify the
equipment used. Do not cover the existing label on the leak detector.
16) Connect the power cord and power up the unit.
17)
Watch the home screen to verify the
Spectube Pressure Wait
message progresses to
Stabilization Wait
and
System Ready
within ten minutes.
Refer to the operator's manual if the system fails to reach the
System Ready
mode.
18) After installation refer to the Leak Detector manual section for leak detector activation of the wireless remote.
19) Refer to the wireless remote user's manual (P/N # 699909945) for operational instructions.
20) Agilent recommends a full calibration of the unit prior to leak test operations.
New Label Location