Figure 5: Antenna Connector
CAUTION
Use proper ESD safety techniques when performing a PCB card install.
8) Connect the antenna cable to the U.FL on the wireless base board (Figure 5: Antenna Connector). Carefully
center the connector on the jack and press the connector onto the jack until it snaps into place. Rotate the
antenna cable to ensure the connector is firmly seated in the jack.
9) Connect the RS-232 on the wireless board using its two slotted screws (Figure 6: RS232 Connection, left side).
Plug in the other side of the ribbon cable to the green port on the CPU board, located on the front side of the leak
detector, away from the antenna's printed circuit board (Figure 6: RS232 Connection, right side).
Figure 6: RS232 Connection
U.FL Jack