Wireless Remote Base Unit for VS Series and G8610, G8611, or G8612 Series Leak Detectors
Page
6
of 14
Document No.
699910002
Revision
D
8) Place the wireless board assembly on the four fan bracket standoffs. Note the correct orientation of the wireless
base board P300 connector (Figure 4: Base Board Assembly with Posts).
9) Connect the antenna cable to the U.FL (Color: silver) on the wireless base board (Figure 5: Antenna Connector).
Carefully center the connector on the jack and press the connector onto the jack until it snaps into place. Rotate
the antenna cable to ensure the connector is firmly seated in the jack.
10) Secure the wireless base board to the fan bracket using the four M3 Phillips screws provided with the kit.
11) Unfurl the black/red wire (P1-RF) with the 24 V three-pin connector and the ground lug (Figure 6: 24 V Power
Cable), which is tie rapped at the base of the leak detector, and:
• Plug the 24 V connector labeled P1-RF into the three-pin connector on the wireless board.
• Attach the 24 V ground lug to one of the bracket screws.
• Reinstall the fan assembly by tightening the M4 socket head cap screw and the two wing nuts (Figure 2: Fan
Assembly and Socket Head Cap Screw).
Figure 6: 24 V Power Cable
12) Connect the RS-232 cable end labeled J300-1 to P300 on the wireless board using its two slotted screws (Figure
5: Antenna Connector).