CAT (Cable and Antenna Test) Mode
21
CAT Mode Settings
Select
CAT Mode
before making any setting in this chapter.
How to select CAT Mode
Press
Mode
.
Then
CAT
.
Measurement Selection
How to select a CAT Mode Measurement
Learn more about the following measurements in the
Supplemental Online
Help:
http://na.tm.agilent.com/fieldfox/help/
Press
Measure 1
.
Then choose one of the following: These softkeys also appear after CAT Mode
is selected.
o
Distance to Fault
1-port reflection measurement that uses Inverse Fourier
Transform (IFT) calculations to determine and display the distance to, and
relative size of, a fault or disruption in the transmission line. Units are in
return loss format, expressed as a positive number in dB, unless the
measurement selected is DTF (VSWR). Learn more about DTF Measurements
on page 31.
o
Return Loss & DTF
Displays both a Return Loss measurement and a DTF
measurement. Use this format to display the frequency settings that are used
to make the DTF measurement. This pair of measurements is created as a
Dual-Trace configuration. Learn more on page 22.
o
Return Loss
1-port reflection measurement that displays the amount of
incident signal energy MINUS the amount of energy that is reflected. The
higher the trace is on the screen, the more energy being reflected back to the
FieldFox. Learn how to measure Return Loss on page 27.
o
VSWR
(Voltage Standing Wave Ratio – also known as SWR) 1-port reflection
measurement that displays the ratio of the maximum reflected voltage over
the minimum reflected voltage. The higher the trace is on the screen, the
more energy being reflected back to the FieldFox.
o
DTF (VSWR)
Distance to Fault in VSWR format.
o
1-Port Cable Loss
1-port reflection measurement that displays the loss of a
transmission line. Learn more on page 27.
o
2-Port Insertion Loss
(Option 110 ONLY) 2-port transmission measurement
that accurately displays the loss through a cable or other device in dB. Both
ends of the cable must be connected to the FieldFox. NO phase information
is included in this measurement. Learn more on page 30.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...