VVM (Vector Voltmeter) Mode – Option 308
129
VVM Calibration
In VVM Mode, calibration is performed in the same manner as in CAT and NA
Modes. In summary, a Preset Cal is sufficient when the DUT is connected
directly to the FieldFox RF Out test port. When using an adapter or jumper cable
to connect the DUT, first perform a Qu Load or OSL Cal before making
measurements. Learn more about these FieldFox calibrations on page 53.
To make an accurate measurement, set
Zero OFF
before calibrating.
VVM Mode Settings
Select VVM Mode before making any setting in this section.
NOTE
In VVM Mode, the power level out of the internal FieldFox source is always
approxi5 dBm and can NOT be changed.
How to select VVM Mode
Press
Mode
Then
VVM
Measurement Selection
How to select a VVM Measurement
Press
Measure 1
Then choose from the following:
o
1-Port Cable Trimming
1-port reflection measurement. Learn how to make
this measurement on page 132.
o
2-Port Transmission
2-port transmission MAGNITUDE ONLY measurement.
Learn how to make this measurement on page 133.
Frequency Selection
VVM measurements are made at a single CW frequency.
How to set Frequency
NOTE
Frequency can be set to 1 Hz resolution.
Press
Freq/Dist
Then
Frequency
Enter a value using the numeric keypad, the
▲|▼
arrows, or the rotary knob.
After using the
▲|▼
arrows or the rotary knob, press
Enter
. The increment
setting of the arrows is based on the current span and can NOT be changed in
VVM Mode.
After using the keypad, select a multiplier key. Learn about multiplier
abbreviations on page 19.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...