NOTE
Although the start frequency can be set as low as -100 MHz, amplitude accuracy
is specified above 2 MHz. Below 2 MHz, frequency accuracy is maintained, but
amplitude accuracy is degraded.
The frequency range of the measurement can be changed using the following
methods:
Set Center and Span frequencies
Set Start and Stop frequencies
Toggle between Zero and Last selected span
Set Full Span of the FieldFox
Using Channel selection. Learn how on page 73.
Using Marker -> Center (frequency). Learn how on page 139.
How to set Frequency Range
Press
Freq/Dist
.
Then choose from the following:
o
Center
and
Freq Span
frequencies. This is the most common method of
entering frequency range. Enter the frequency of interest. The Center
frequency is at the exact center of the X-axis. The Frequency Span places
half of the frequency range on either side of center. When the Center and
Frequency Span values are entered, then the X-axis annotation on the screen
shows the Center and Span frequencies.
o
Start
and
Stop
frequencies. Start is the beginning of the X-axis and Stop is
the end of the X-axis. When the Start and Stop frequencies are entered, then
the X-axis annotation on the screen shows the Start and Stop frequencies.
Follow each of these by entering a value using the numeric keypad, the
▲|▼
arrows, or the rotary knob.
When using the
▲|▼
arrows, press
Enter
. The increment setting of the arrows
is based on the current span. This can be changed in SA Mode. See
How to
change frequency step size
below.
After using the keypad, select a multiplier key. Learn about multiplier
abbreviations on page 19.
Two more ways to change the frequency range
Press
Freq/Dist
.
Then choose from the following:
o
Zero/Last Span
Each key press toggles the frequency span between 0 Hz
and the last specified frequency span.
o
More
o
Then
Full Span
Selects the entire frequency span of the FieldFox. The
Center frequency is set automatically.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...