Chapter 2
45
Basic Operation
Configuring the RF Output
Configuring a Ramp Sweep for a Master/Slave Setup
This procedure shows you how to configure two PSGs and an 8757D to work in a master/slave setup.
1. Set up the equipment as shown in
Figure 2-7
. Use a 9-pin, D-subminiature, male RS-232 cable with the
pin configuration shown in
Figure 2-8 on page 46
to connect the auxiliary interfaces of the two PSGs.
You can also order the cable (part number 8120-8806) from Agilent Technologies.
By connecting the master PSG’s 10 MHz reference standard to the slave PSG’s 10 MHz reference input,
the master’s timebase supplies the frequency reference for both PSGs.
Figure 2-7
Master/Slave Equipment Setup
Summary of Contents for E8247C
Page 10: ...Contents x ...
Page 96: ...86 Chapter 4 Analog Modulation Configuring the LF Output ...
Page 142: ...132 Chapter 6 Custom Arb Waveform Generator Working with Filters ...
Page 178: ...168 Chapter 7 Custom Real Time I Q Baseband Working with Differential Data Encoding ...
Page 198: ...188 Chapter 10 Troubleshooting ...
Page 214: ...204 Chapter 10 Troubleshooting Returning a Signal Generator to Agilent Technologies ...