Chapter 2
39
Basic Operation
Configuring the RF Output
6. Preset either instrument.
Presetting one of the instruments should automatically preset the other as well. If both instruments do
not preset, check the GPIB connection, GPIB addresses, and ensure the 8757D is set to system interface
mode (
SYSINTF
set to ON).
The PSG automatically activates a 2 GHz to maximum frequency ramp sweep with a constant amplitude
of 0 dBm. Notice that the
RF ON
,
SWEEP
, and
PULSE
annunciators appear on the PSG display. The
PULSE
annunciator appears because the 8757D is operating in AC mode.
The PSG also switches its remote language setting to 8757D System, allowing the PSG to talk to the
8757D during ramp sweep operations. You can confirm this by pressing
Utility
>
GPIB/RS-232 LAN
and
observing the selection under the
Remote Language
softkey.
NOTE
During swept RF output, the
FREQUENCY
and/or
AMPLITUDE
areas of the signal
generator’s display are deactivated, depending on what is being swept. In this case,
since frequency is being swept, nothing appears in the
FREQUENCY
area of the display.
7. Press
Frequency
>
Freq CW
.
The current continuous wave frequency setting now controls the RF output and ramp sweep is turned off.
8. Press
Freq Start
.
The ramp sweep settings once again control the RF output and the CW mode is turned off. Pressing any
one of the softkeys
Freq Start
,
Freq Stop
,
Freq Center
, or
Freq Span
activates a ramp sweep with the current
settings.
NOTE
In a frequency ramp sweep, the start frequency must be lower than the stop frequency.
9. Adjust the settings for
Freq Center
and
Freq Span
so that the frequency response of the device under test
(DUT) is clearly seen on the 8757D display.
Notice how adjusting these settings also changes the settings for the
Freq Start
and
Freq Stop
softkeys. You
may need to rescale the response on the 8757D for a more accurate evaluation of the amplitude.
Figure
2-3 on page 40
shows an example of a bandpass filter response.
Summary of Contents for E8247C
Page 10: ...Contents x ...
Page 96: ...86 Chapter 4 Analog Modulation Configuring the LF Output ...
Page 142: ...132 Chapter 6 Custom Arb Waveform Generator Working with Filters ...
Page 178: ...168 Chapter 7 Custom Real Time I Q Baseband Working with Differential Data Encoding ...
Page 198: ...188 Chapter 10 Troubleshooting ...
Page 214: ...204 Chapter 10 Troubleshooting Returning a Signal Generator to Agilent Technologies ...