Configuring Hardware
•
“To Set the ARB Reference”
see
page 144
To Set a Delayed, Positive Polarity, External Single Trigger
Using this procedure, you learn how to utilize an external function generator to apply a delayed
single-trigger to a custom multicarrier waveform.
1. Connect an Agilent 33120A function generator or equivalent to the signal generator PATT TRIGGER IN
port, as shown in
Figure 6-1
.
Figure 6-1
2. On the signal generator, press
Preset
.
3. Press
Mode
>
Custom
>
Arb Waveform Generator
.
4. Press
Multicarrier Off On
until On is highlighted.
5. Press
Trigger
>
Single
.
6. Press
Trigger
>
Trigger Setup
>
Trigger Source
>
Ext
.
7. Press
Ext Polarity Neg Pos
until Pos is highlighted.
8. Press
Ext Delay Off On
until On is highlighted.
9. Press
Ext Delay Time
>
100
>
msec
.
The Custom Arb Waveform Generator has been configured to play a single multicarrier waveform
100 milliseconds after it detects a change in TTL state from low to high at the PATT TRIG IN rear panel
connector.
10. Set the function generator waveform to a 0.1 Hz square wave at an output level of 0 to 5V.
Summary of Contents for E8247C
Page 10: ...Contents x ...
Page 96: ...86 Chapter 4 Analog Modulation Configuring the LF Output ...
Page 142: ...132 Chapter 6 Custom Arb Waveform Generator Working with Filters ...
Page 178: ...168 Chapter 7 Custom Real Time I Q Baseband Working with Differential Data Encoding ...
Page 198: ...188 Chapter 10 Troubleshooting ...
Page 214: ...204 Chapter 10 Troubleshooting Returning a Signal Generator to Agilent Technologies ...