820
W_t=0
!
Specifies
the
default
wait
time
per
!
frequency.
830
INPUT
"Search
mode?[1-6]",Search_mode
!
Inputs
the
search
mode
(intervals
!
at
which
the
frequency
changes).
840
INPUT
"Waiting
time?[ms]",W_t1
!
Inputs
the
wait
time
in
[msec].
850
W_t=W_t1*1.E-6
!
Converts
the
wait
time
to
that
in
[sec].
860
!
870
RETURN
880
!
890
Pi_cal:
!
Starts
the
fixture
correction
routine.
900
OUTPUT
@E5100;";CALI
ONEP"
!
Specifies
the
1-port
3-term
!
correction
method.
910
INPUT
"Leave
the
terminal
OPEN,
and
press[return].",Cal$
!
Instructs
that
!
the
terminal
be
left
open.
920
OUTPUT
@E5100;";CLASS11A"
!
Selects
OPEN
for
the
S11
correction
!
standard
class.
930
INPUT
"Connect
SHORT,
and
press[return]",Cal$
!
Instructs
that
the
!
terminal
be
shorted.
940
OUTPUT
@E5100;";CLASS11B"
!
Selects
SHORT
for
the
S11
!
correction
standard
class.
950
INPUT
"Connect
LOAD,
and
press[return]",Cal$
!
Instructs
that
the
load
!
be
connected
to
the
terminal.
960
OUTPUT
@E5100;";CLASS11C"
!
Selects
LOAD
for
the
S11
correction
!
standard
class.
970
OUTPUT
@E5100;";SAV1"
!
Calculates
and
stores
the
error
facto
!
for
the
3-term
correction.
980
RETURN
990
END
Measuring
Load
Resonance
Frequency
(FL)
Using
a
CL
Board
with
Fixed
C
K
-5
Summary of Contents for E5100A
Page 1: ...Agilent E5100A B Network Analyzer Programming Manual ...
Page 2: ......
Page 5: ...c Copyright 1995 1996 1997 1998 2000 2001 Agilent Technologies Japan Ltd ...
Page 8: ......
Page 36: ......
Page 46: ......
Page 48: ...Figure 3 1 Simpli ed Data Processing Flow 1 3 2 Data Processing and T ransfer ...
Page 58: ......
Page 76: ......
Page 112: ......
Page 232: ......
Page 246: ......
Page 250: ......
Page 299: ...110 END Waveform Analysis Commands D 49 ...
Page 302: ......
Page 322: ......
Page 350: ......
Page 352: ......
Page 362: ......
Page 368: ......
Page 372: ......