6.
Returns
the
maximum
height
of
the
ripple
,
Rpl
3
,
that
is
the
dierence
between
the
peak
and
left
adjacent
negative
peak.
Note
Y
ou
must
select
the
following
conditions
to
use
this
command:
Dual
Channel
&
Coupled
Channel:
ON
Impedance
Conversion:
ON
Analysis
channel:
LOG
MA
G
format
Non-analysis
channel:
Phase
format
OUTPRESR?
returns
the
rst
two
0
phase
point
events
found
if
there
are
more
than
three
corresponding
points
points
.
If
there
is
only
one
0
phase
point
in
the
range
,
OUTPRESR?
denes
that
point
as
a
resonant
point
and
returns
0
for
Z
a
,
f
a
,
Rpl
1
,
Rpl
2
,
and
Rpl
3
.
If
there
is
no
0
point,
OUTPRESR?
returns
0
for
all
parameters
.
Examples
10
ASSIGN
@E5100
TO
717
20
OUTPUT
@E5100;"DUAC
ON;
COUC
ON"
30
OUTPUT
@E5100;"CHAN2;
FMT
PHAS"
40
OUTPUT
@E5100;"CHAN1;
FMT
LOGM;
CONV
ZTRA"
50
OUTPUT
@E5100;"CENT
70MHZ;
SPAN
100KHZ"
60
CALL
Sweep(2)
!
Goes
to
sub
routine.
(See
OUTPFILT?)
70
OUTPUT
@E5100;"ANAOCH1;ANARFULL;ANAODATA"
80
OUTPUT
@E5100;"OUTPRESR?"
90
ENTER
@E5100;Zr,Fr,Za,Fa,R1,R2,R3
100
"Resonant:";Zr;"[ohm],";Fr;"[Hz]"
110
"Anti-Resonant:";Za;"[ohm],";Fa;"[Hz]"
120
"Ripple
L:";R1;"[dB]"
130
"Ripple
M:";R2;"[dB]"
140
"Ripple
R:";R3;"[dB]"
150
END
Waveform
Analysis
Commands
D-39
Summary of Contents for E5100A
Page 1: ...Agilent E5100A B Network Analyzer Programming Manual ...
Page 2: ......
Page 5: ...c Copyright 1995 1996 1997 1998 2000 2001 Agilent Technologies Japan Ltd ...
Page 8: ......
Page 36: ......
Page 46: ......
Page 48: ...Figure 3 1 Simpli ed Data Processing Flow 1 3 2 Data Processing and T ransfer ...
Page 58: ......
Page 76: ......
Page 112: ......
Page 232: ......
Page 246: ......
Page 250: ......
Page 299: ...110 END Waveform Analysis Commands D 49 ...
Page 302: ......
Page 322: ......
Page 350: ......
Page 352: ......
Page 362: ......
Page 368: ......
Page 372: ......