Chapter 2
37
Performance Test
7. DYNAMIC ACCURACY TEST
2. Pe
rfor
ma
nce T
est
7. DYNAMIC ACCURACY TEST
Description
This test checks the dynamic accuracy of the E5071C. The dynamic accuracy is tested at
1.195 GHz for each receiver port using the Agilent Z5623A (with option H01) Dynamic
Accuracy Test Kit. The 1.195 GHz source signal of the E5071C goes through the Z5623A's
first attenuator variable from 0 dB to 11 dB in 1 dB steps and second attenuator variable up
to 100 dB in 10 dB steps and is measured with the receiver port. The test procedure checks
the measured receiver-input power for each 5 dB increment in the range from 10 dBm to
100 dBm. The ratios of the measured receiver-input powers to the reference input level of
-10 dBm are calculated and compared to the attenuation values of the Z5623A. The 1.195
GHz signal flow is split with a power splitter present between the two attenuators in the
Z5623A and measured with an external power meter to calibrate the attenuated power
level.
NOTE
Since the dynamic accuracy does not have frequency dependence, this test is performed at
1.195 GHz only.
Test Equipment
Power meter
Agilent E4419A/B
Power sensor
Agilent 8482A
Dynamic Accuracy Test Kit
Agilent Z5623A w/Opt. H01
6 dB fixed attenuator
Agilent 8491A w/Opt. 006 and
H60
Coaxial cable with N-type (m) connectors, 61 cm
(24 in), 2 ea.
Agilent N6314A (p/n 8120-8862)
GPIB controller (or PC with GPIB Interface Card)
GPIB Interconnection Cable
Agilent 10833A/B
NOTE
Before starting this test, input and store the calibration data of 8482A to E4419A/B.
Summary of Contents for E5071C
Page 14: ...14 Contents ...
Page 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Page 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Page 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Page 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Page 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Page 364: ...364 AppendixC Firmware Update Firmware Update ...
Page 365: ...365 D Power Requirement D Power Requirement ...