Chapter 2
137
Performance Test
E5071C Performance Test Record
2. Pe
rfor
ma
nce T
est
Dynamic Accuracy Test
@ Reference power level: -10 dBm
Direction: S21
Power level [dBm]
Test limit [dB]
Test result [dB]
Measurement
uncertainty [dB]
10
±
0.207
±
0.010
5
±
0.075
±
0.010
0
±
0.041
±
0.007
-5
±
0.031
±
0.009
-15
±
0.030
±
0.006
-20
±
0.035
±
0.007
-25
±
0.040
±
0.010
-30
±
0.045
±
0.011
-35
±
0.056
±
0.015
-40
±
0.067
±
0.019
-45
±
0.078
±
0.021
-50
±
0.091
±
0.021
-55
±
0.106
±
0.024
-60
±
0.125
±
0.024
-65
±
0.151
±
0.026
-70
±
0.189
±
0.025
-75
±
0.248
±
0.011
-80
±
0.346
±
0.014
-85
±
0.509
±
0.021
-90
±
0.785
±
0.033
-95
±
1.248
±
0.069
-100
±
2.008
±
0.060
Direction: S12
Power level [dBm]
Test limit [dB]
Test result [dB]
Measurement
uncertainty [dB]
10
±
0.207
±
0.010
5
±
0.075
±
0.010
Summary of Contents for E5071C
Page 14: ...14 Contents ...
Page 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Page 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Page 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Page 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Page 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Page 364: ...364 AppendixC Firmware Update Firmware Update ...
Page 365: ...365 D Power Requirement D Power Requirement ...