Chapter 2
31
Performance Test
2. FREQUENCY ACCURACY TEST
2. Pe
rfor
ma
nce T
est
2. FREQUENCY ACCURACY TEST
Description
This test checks the frequency accuracy of the E5071C test port output signal. The
frequency accuracy is checked at 50 MHz and 8.5 GHz with a frequency counter. Since the
E5071C employs a PLL frequency synthesizer for the signal source, the frequency
accuracy test at these two frequency points can verify the accuracy for the entire frequency
range.
Test equipment
Frequency Counter
Agilent 53181A with Opt. 010 and 124 or Agilent
53131/2A with Opt. 010 and 124
48 inch BNC cable
Agilent p/n 8120-1840
N(m)-BNC(f) adapter
Agilent p/n 1250-0780