Chapter 4
177
Troubleshooting
How to exit from the E5071C Measurement View
4. T
ro
uble
shoo
tin
g
How to exit from the E5071C Measurement View
You need to exit from the E5071C Measurement View to perform some troubleshooting.
The following is the procedure to exit from the E5071C Measurement View.
Step 1.
Connect the mouse and external keyboard to the connectors on the E5071C rear panel.
Step 2.
Turn the instrument on.
Step 3.
Press
key.
Step 4.
Click
Service Menu
-
Service Function
. Password dialog box opens as shown in Figure
4-1.
Figure 4-1
Password dialog box
e5071cse1057
Step 5.
Enter the password
kid
in the Password box and then click "OK" button.
Step 6.
Click
Exit
in Service Functions Menu. Then the E5071C exit the Measurement View,
then.windows desktop screen appears with 3 icons (Network Analyzer, Recycle Bin, and
Internet Explorer.)
NOTE
If you wish to return to the Measurement View, double-click "Network Analyzer" icon.
NOTE
If you need to shut down the E5071C and again turn on, perform in accordance with the
following procedure.
a.
To get "Start" menu bar displayed, move the pointer to the bottom of the screen with
mouse.
b.
Click "Start" and "Turn off Computer" in the pull down menu. "Turn off Computer"
dialog box opens.
c.
Click "Turn off" button in the dialog box.
Summary of Contents for E5071C
Page 14: ...14 Contents ...
Page 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Page 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Page 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Page 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Page 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Page 364: ...364 AppendixC Firmware Update Firmware Update ...
Page 365: ...365 D Power Requirement D Power Requirement ...