Programming Reference
Detailed Command Descriptions
118
E1852B Operating Guide
CONFigure:RF:ATT:RX <value>
Control RF attenuator in the receive direction
CONFigure:RF:DIRTY_TX <value>
Set Impairments
Syntax:
CONFigure:RF:ATT:RX <value>
Value Range:
ON
Receive path Attenuation
(25 dB)
Default Value:
OFF
OFF
Zero path attenuation.
Applicable in
State:
Set: All
Query: All
Description:
Use this command to add attenuation in the receive signal path by
turning the 25 dB attenuator on. Sending the switch ‘OFF’ sets the
attenuation to 0 dB.
Note: 0dB attenuation increases the input impedance of test set.
Examples:
CONF:RF:ATT:RX ON
Places 25 dB
attenuation in
the receive signal
path
CONF:RF:ATT:RX?
Returns current
setting
Syntax:
CONFigure:RF:DIRTY_TX <value>
Value Range:
ON
Enable dirty transmitter mode Default Value:
OFF
OFF
Disable dirty transmitter.
Applicable in
State:
Set: All
Query: All
Description:
Use this command to add the dirty transmitter impairments
RF:DIRTY_TX:OFFSet
and
RF:TXMODulation:INDEX
to the transmitted signal.
Examples:
CONF:RF:ATT:DIRTY_TX ON
Switches dirty
transmitter mode
on
CONF:RF:ATT:DIRTY_TX?
Returns current
setting
E1852-90007.book Page 118 Monday, February 4, 2002 2:09 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com