PC User Interface
System Page
32
E1852B Operating Guide
Table 1
System Page - Panel Summary
Panel
Parameters
Description
1. Communication
Port
Lpt1
Lpt2
Lpt3
PC parallel ports for control of the test set.
Lpt1 is default
2. GPIB
Configuration
GPIB card
GPIB address
Used to configure the GPIB system for
communication with the test set
3. Firmware Update
Com1 to Com 9
Used to download new firmware to the
test set
4. Wait in Testmode
Setup
Enabled/Disabled
The test set pauses during a testmode
setup as required by some DUTs
5. Device
Features
Readback
enabled/disabled
Displays the DUT enabled Bluetooth
features
6. Windows Position Default
Pressing Default places the user interface
in the top left corner of your PC desktop
7. Hardware
Reset
Pressing Reset initiates a ’warm start’
reset of the test set
8. Software Version
Firmware Version
The test set is polled and the Firmware
version is displayed
PC DLL Version
Displays the DLL version held on the PC
9. Log Settings
Log Measurement
Log file containing all the measurements
Log Interval (sec.)
Length of time between data logging
events in seconds
Log Communication
In Debug mode only, log file containing all
the measurements and commands
10. Tester and
DUT address
Tester BD Address
Entry and display of the BD address of the
test set
DUT BD Address
Entry and display of the BD address of the
DUT
11. Address Inquiry
DUT BD Addresses
Used to find out addresses of any BD in
the vicinity of the test set
E1852-90007.book Page 32 Monday, February 4, 2002 2:09 PM