Programming Reference
Detailed Command Descriptions
126
E1852B Operating Guide
CONFigure:SLAVe:BDADdr <value>
Slave Bluetooth Device Address (DUT)
CONFigure:TESTmode:MODE <value>
Testmode Mode selection
Syntax:
CONFigure:SLAVe:BDADdr <value>
Value Range:
12 digits in hex format
Default Value:
000000000020
Applicable in
State:
Set: All except CONN, and TCONN
Query: All
Description:
This command configures the DUT address in the test set
Examples:
CONF:SLAV:BDAD AFC2DE56312F
Sets address
CONF:SLAV:BDAD?
Returns current
setting
Syntax:
CONFigure:TESTmode:MODE <value>
Value Range:
LOOPback
Loopback test mode
Default Value:
TRAN
TRANsmit
Transmitter test mode
Applicable in
State:
Set: All
Query: All
Description:
This command configures either transmitter or loopback test mode
operation of the test set and DUT.
Examples:
CONF:TEST:MODE LOOP
Sets loopback
test mode
CONF:TEST:MODE?
Returns current
setting
E1852-90007.book Page 126 Monday, February 4, 2002 2:09 PM