PC User Interface
Set-up Page
66
E1852B Operating Guide
How the Limits are used
Four limits are required for each parameter.
Min
,
Min OK
,
Max OK
,
and
Max
are used to scale the red/green bars for each of the associated
parameters on the
Measurements
windows. (see Figure 41 on Page 66)
–
Min
and
Max
limits set the end points of the bar.
–
Min OK
and
Max OK
limits set the position and size of the green
‘OK’ or ‘Pass’ section.
– Sections of the bar between the
Min OK
and
Max OK
regions are
colored red to indicate a ‘Fail’.
– The measurement result is shown numerically in the display field.
It is also indicated along the length of the bar by a marker.
– The background area is red when the result is outside the ‘OK’
limits and changes to green when within the ‘OK’ limits. (see
Figure 41 and Figure 42 on Page 67)
Figure 41
Frequency Drift Limits and Display Bar (Pass)
Indicator
(Pass)