460
Agilent E5505A User’s Guide
21
Preventive Maintenance
When reconnecting this type of cable:
•
Carefully insert the male connector center pin into the female connector.
(Make sure the cable is aligned with the instrument connector properly
before joining them.)
•
Turn the silver nut clockwise by hand until it is snug, then tighten with an 8
inch-lb torque wrench (part number 8720-1765).
Bent Semirigid Cables
Semirigid cables are not intended to be bent outside of the factory. An
accidental bend that is slight or gradual may be straightened carefully by hand.
Semirigid cables that are crimped will affect system performance and must be
replaced. Do not attempt to straighten a crimped semirigid cable.