Residual Measurement Examples
8
Agilent E5505A User’s Guide
233
14
When you have completed these operations, click the
Close
button.
Setup considerations for amplifier measurement
Connecting cables
The best results will be obtained if semi-rigid coaxial cables are used to
connect the components used in the measurement; however, BNC cables have
been specified because they are more widely available. Using BNC cables may
degrade the close-in phase noise results and, while adequate for this example,
should not be used for an actual measurement on an unknown device unless
absolutely necessary.
Measurement environment
The low noise floors typical of these devices may require that special attention
be given to the measurement environment. The following precautions will help
ensure reliable test results:
•
Filtering on power supply lines
•
Protection from microphonics
•
Shielding from air currents may be necessary.
Figure 168
Select graph description on graph tab
e5505a_user_select_graph_desc_tab
24 Jun 04 rev 3
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...